Articles in this section

TotalColor - Slow Behavior

Symptom:

TotalColor slows down on certain measurements.

Solution:

1) In up/down mode, the Spectro LFP increases the measuring time when measuring dark colors. This increases the measurement accuracy, but also the measurement time.
For sensitive or sticky materials we recommend to use the contact less mode. In this mode, the measuring head does not touch the surface of the chart.

2) In case of transmission measurements, the measuring time depends on the density of the material you measure.
As higher the density of the material as higher the measuring time.

 

3) Change the Transmission to Reflection to see if it increases the speed

Screenshot_2024-05-06_at_5_09_50 PM.png

4) Increase/reduce the calibration frequency.

 

Was this article helpful?
0 out of 0 found this helpful