TotalColor LFP - No measurement head/arm up or down movement

Symptom:

After starting the measurement, the "head/arm" doesn't move up or down.

Problem:

The measuring head/arm should move down in order to determine the thickness of the target when starting the measurement sequence.

Solution:

The problem might be caused by an inactive micro-switch.

Switch the instrument off and on again to ensure that the device moves the measurement head/arm to its initial position.

Remove the cap by unscrewing the 4 screws on the left and the right side as indicated in the picture:

Screen_Shot_2019-05-31_at_11.46.45_AM.png

Adjust the screw marked by the arrow until the micro-switch is active (you can hear when the switch gets activated).

It typically only requires half a turn.

Screen_Shot_2019-05-31_at_11.48.20_AM.png

Remount the cap and fix it with the 4 screws.
 
After this adjustment should the measuring head should move down in order to recognize the thickness of the target when starting a measurement.

 