The information below corresponds to an issue found in CalderaRIP 14.0, and other previous versions.
However, the issue was fixed in CalderaRIP V15.1 (see changelog here), as starting from this version there are two separate drivers for each device.
Issue
Reading error #2110 when reading a regular target with the i1Pro3 Plus with the IO table.
Problem
The current i1IO3 driver supports both i1IO3 with regular targets and the i1IO3 Plus with large aperture targets, however, the i1Pro3 Plus doesn't support the regular (read smaller targets) targets but only the large aperture targets.
Solution
For users of CalderaRIP V15.1 or after:
Use one of the two available drivers, depending on your needs.
Starting with this version, the X-Rite i1 iO3 driver manages the standard aperture targets, and the X-Rite i1 iO3 Plus driver manages the large aperture targets.
For users of CalderaRIP V15 or before:
When using the i1Pro3 Plus in the IO table, you need to use the large aperture targets: